User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: PM 2502 VERIFICATION PROCEDURE (Ver1.0) DATE: 03-Oct-93 AUTHOR: User Contributed REVISION: 1.0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 35 NUMBER OF LINES: 90 CONFIGURATION: Fluke 5700A ============================================================================= * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * Source: Service Manual PM P/N: @ * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * TUR Configuration: * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * When the T.U.R. is < 2:1, the source instrument, e.g. 5700A, must be * characterized and appropriate corrections entered in the correction file * (e.g. 5700.COR). Where T.U.R.'s are > 2:1 but < 4:1, the test tolerance * is reduced by root sum square (rss) techniques. * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * This procedure is not executable in MET/CAL versions below ver. 2.00 * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * Fluke make no warranty, expressed or implied, as to the fitness or * useability of this procedure. * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * Main Procedure TESTS: Verification * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * The following subprocedures are called from this main procedure: * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R P M F L W 1.002 DISP ! CHECK THE POSITIVE AND NEGATIVE BATTERIES BY 1.002 DISP ! SELECTING THE APPROPRIATE FUNCTION BEFORE 1.002 DISP ! VERIFICATION OF THE INSTRUMENT 1.002 DISP ! THEN CHECK ZERO IN DC-V AS WELL AS IN AC-V 1.003 HEAD 1.004 DISP [27][91]1m Connect as follows 1.004 DISP ! ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ 1.004 DISP ! [27][91]1m 5700 UUT 1.004 DISP ! Sense HI ÄÄÄÄÄÄÄÄÄ¿ 1.004 DISP ! Output HI ÄÄÄÄÄÄÄÄÄÁÄ> Volt/Ohm Input 1.004 DISP ! Output LO ÄÄÄÄÄÄÄÄÄÂÄ> Common 1.004 DISP ! Sense LO ÄÄÄÄÄÄÄÄÄÙ 1.005 HEAD { DC VOLTAGE PERFORMANCE VERIFICATION} 1.006 DISP SELECT THE +DC VOLTS FUNCTION 1.007 5700 100mV 3.00% 2W 2.001 5700 300mV 1.50% 2W 3.001 5700 1V 1.50% 2W 4.001 5700 3V 1.5% 2W 5.001 5700 10V 1.5% 2W 6.001 DISP SELECT THE -DC VOLTS FUNCTION 6.002 5700 -10V 1.5% 2W 7.001 DISP SELECT THE +DC VOLTS FUNCTION 7.002 5700 30V 1.5% 2W 8.001 5700 100V 1.5% 2W 9.001 5700 300V 1.5% 2W 10.001 5700 1000V 1.5% 2W 11.001 HEAD {ALTERNATING VOLTAGE PERFORMANCE VERIFICATION} 11.002 DISP SELECT THE AC-VOLTS FUNCTION 11.003 5700 1V 2.5% 60H 2W 12.001 5700 3V 2.5% 60H 2W 13.001 5700 10V 2.5% 60H 2W 14.001 5700 10V 5% 3kH 2W 15.001 5700 30V 2.5% 60H 2W 16.001 5700 100V 2.5% 60H 2W 17.001 5700 300V 2.5% 60H 2W 18.001 5700 1000V 2.5% 60H 2W 19.001 HEAD {RESISTANCE PERFORMANCE VERIFICATION} 19.002 DISP SELECT THE BUZZER AND AC FUNCTION 19.003 5700 1Z S 4W 19.004 DISP SELECT THE +DC VOLTS FUNCTION AND *1 OHM RANGE 19.005 HEAD {ADJUST *1 OHM RANGE TO 0 OHM} 19.006 5700 0Z 99% 2W 20.001 HEAD {*1 OHM VERIFICATION} 20.002 5700 19Z 8% 4W 21.001 HEAD {ADJUST *100 OHM RANGE TO 0 OHM} 21.002 5700 0Z TOL 2W 22.001 HEAD {*100 OHM VERIFICATION} 22.002 5700 1900Z 5% 4W 23.001 HEAD {ADJUST *10 KOHM RANGE TO 0 OHM} 23.002 5700 0Z TOL 2W 24.001 HEAD {*10 KOHM VERIFICATION} 24.002 5700 190kZ 5% 2W 25.001 HEAD {DC CURRENT PERFORMANCE VERIFICATION} 25.002 DISP SELECT +DC FUNCTION 100uA RANGE 25.003 5700 100uA 2.5% 2W 26.001 5700 1mA 2.5% 2W 27.001 5700 10mA 2.5% 2W 28.001 5700 100mA 2.5% 2W 29.001 5700 1A 2.5% 2W 30.001 HEAD {AC CURRENT PERFORMANCE VERIFICATION} 30.002 DISP SELECT AC FUNCTION 30.003 5700 100uA 2.5% 60H 2W 31.001 5700 1mA 2.5% 60H 2W 32.001 5700 10mA 2.5% 60H 2W 33.001 5700 100mA 2.5% 60H 2W 34.001 5700 1A 2.5% 60H 2W 35.001 HEAD 35.002 DISP [27][91]1m PLEASE SWITCH THE PM2502 FROM THE AC FUNCTION 35.002 DISP [27][91]1m TO THE OHMS DC FUNCTION, PREFERABLY WITH THE 35.002 DISP [27][91]1m DIAL IN A HIGH VOLTAGE RANGE 35.002 DISP [D1000][007][D800][007][D600][007][D500][007] 35.002 DISP [D400][007][D300][007][D200][007][D100][007] 35.002 DISP [D80][007][D60][007][D50][007][D40][007][D30][007] 35.003 DISP ! P R O C E D U R E C O M P L E T E 35.003 DISP ! Disconnect all cables 35.003 DISP ! [27][91]7m Complete required paperwork 35.004 END